EasyManuals Logo

AVL DiGas 4000 User Manual

Default Icon
182 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #24 background imageLoading...
Page #24 background image
Error Messages and Troubleshooting AVL 4000
2-2 Service Manual
2.1.2 System Needs Restarting
If two consecutive attempts at restarting do not return the tester to normal operating
mode, the measuring instruments unit must be defective.
Remove the unit (see Dismantling the 4-/5-Gas Measuring Instruments Unit) and
check the
measuring cell
receiver
light source
Check the cable between the main board and the measuring instruments unit
2.1.3 Measuring Chamber Temperature Too High
Remove the probe from the exhaust pipe (because when this error occurs, the tester
switches to zero calibration mode and needs to draw in fresh air).
Do not switch the tester off!
Wait at least 10 min
Acknowledge the error by pressing >>,
Select “Measurement” again.
If the same error message appears again, there is a fault in the measuring instruments
unit.
Remove the unit (see Dismantling the 4-/5-Gas Measuring Instruments Unit).
Check the cable to the receiver / replace the unit if necessary (i.e. filter/receiver).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the AVL DiGas 4000 and is the answer not in the manual?

AVL DiGas 4000 Specifications

General IconGeneral
BrandAVL
ModelDiGas 4000
CategoryTest Equipment
LanguageEnglish

Related product manuals