Periodic Maintenance AVL 4000
1-6 Service Manual
1.3.3 AVL DiCom 4000
Evaluation unit
Check using software
Raw data from 4-gas measuring instruments unit OK?
O
2
sensor is OK?
NO sensor is OK?
Visual inspection
Enter in maintenance log book
Cleaning
Replace particulate filter
(see section 5.3.3 Particulate Filter)
Check O-ring paper filter
Visual inspection of probe and hose
Opacity measuring chamber type A
Visual check
Silicon seal
Leaks in housing
Housing components (cracks, faults)
Closure fittings, snap mechanism of optical elements
Electrical plug connections and cables
Probe connection plate O-ring
Probe hose, probe clamp
Soiling of measuring chamber, purging air channel
Function check
(each test is described in the Operating Manual)
Test function of fans
Test function of heating
Check changeover valve
("Calibration", "Test", "Deactivate" settings - visual check or diesel service mode)
Clean the optical elements (by wiping)
Linear check