4000 Periodic Maintenance
Service Manual 1-11
Opacity measuring chamber type 4000
Visual check
Silicon seal
Leaks in housing
Housing components (cracks, faults)
Closure fittings, snap mechanism of optical elements
Electrical plug connections and cables
Probe hose, probe clamp
Soiling of measuring chamber, purging air channel
Function check
(each test is described in the Operating Manual)
Test function of fans
Test function of heating
Check changeover valve
Clean the optical elements (by wiping)
Linear check
Cleaning
Replace particulate filter
(see section 5.3.3 Particulate Filter)
Check O-ring paper filter
Visual inspection probe and hose
Clean condensate pump
(using cleaning kit)
Function test
Leak test
Gas calibration