Measurements
Damping Factor
[5 s]
U
Use Damping Factor to establish the stability of the measured flow rate. If back and forth oscillations
of the flow rate are observed during normal flow conditions, increase this value incrementally until the
flow rate stabilizes. This parameter has no affect on the totalizers.
To change the Damping Factor value, follow these steps from the Measurements menu.
1. Select Damping Factor to view the Damping Factor display.
2. Select one of the following damping factors:
1 Second 10 Seconds
2 Seconds 20 Seconds
3 Seconds 30 Seconds
4 Seconds No Damping
5 Seconds
3. Press [E] to save the damping factor and return to the Measurements menu.
Inputs/Outputs
Analog Output Range
[4 to 20 mA]
S
Use Analog Output to establish the range of the analog output signal. To change
Analog Output range, follow these steps from the Inputs/Outputs menu:
1. Select Analog Output to view the Analog Output display.
2. Select one of the following options:
4 to 20 mA
0 to 20 mA
2 to 10 mA
0 to 10 mA
3. Press [E] to save the analog output and return to the Inputs/Outputs menu.
OTE: N If an error message is displayed, set current to 22 mA. If you select bidirectional
operation, you can signal flow direction via digital outputs.
Alarm Mode
[OFF]
S
Use Alarm Mode to configure the behavior of the analog output during alarm
conditions. There are three options for this parameter: OFF, LOW, and HIGH.
OFF: Analog signal is based on flow rate and always within the configured range
LOW: During alarm conditions, the analog signal is 2 mA less than the configured
lower range
HIGH: During alarm conditions, the analog signal is 2 mA more than the configured
upper range
For example, if the analog range is 4…20 mA and the alarm mode is set to HIGH, then
during a Full Scale Flow alarm condition, the analog output current is 22 mA.
To change the analog output alarm mode, follow these steps from the Inputs/Outputs
menu:
1. Select Alarm Mode to view the Alarm Mode display.
2. Select one of the following options:
OFF
LOW
HIGH
3. Press [E] to save the alarm mode and return to the Inputs/Outputs menu.
Using the Main Menu Programming Options
Page 40 March 2022MAG-UM-01272-EN-21