Inputs/Outputs
Digital Output
(continued)
Full Scale
Frequency
[1000 Hz]
S
Use Full Scale Frequency to establish the Full Scale Flow output frequency when the
flow rate equals the configured Full Scale Flow.
To change the Full Scale Frequency, follow these steps from the Inputs/Outputs
mainmenu:
1. Select Digital Output 1, 2 or 3 and press [E] to open the Digital Output menu.
2. From the Digital Output menu select Full Scale Frequency and press [E] to
open the Full Scale Frequency display.
3. Select a value for full scale frequency.
4. Press [E] to save the new parameter and return to the Digital Output menu.
Preset Amount
[0.0]
S
Use Preset Amount to set the reset value for the associated PS totalizer when the
digital input is set to Batch Reset.
To change the Preset Amount, follow these steps from the Inputs/Outputs menu:
1. Select Digital Output 1, 2, 3 or 4 and press [E] to open the Digital
Output menu.
2. From the Digital Output menu select Preset Amount and press [E] to open the
Preset Amount display.
3. Enter the preset amount value. Press [E] to save the new parameter and return
to the Digital Output menu.
OTE: N You can only set one Preset Amount. If you set the Preset Amount for Digital
Output 1, it is the same for 2, 3 and 4.
Set Point
Minimum
[0%]
S
Use Set Point Minimum to establish, as a percentage of Full Scale Flow, the threshold
at which the output alarm is activated. Flow rates below the threshold activate the
outputalarm.
To change the Set Point Minimum, follow these steps from the Inputs/Outputs menu:
1. Select Digital Output 1, 2, 3 or 4 and press [E] to open the Digital
Output menu.
2. From the Digital Output menu select Set Point Minimum and press [E] to open
the Set Point Minimum display.
3. Enter the set point minimum value. Press [E] to save the new parameter and
return to the Digital Output menu.
Set Point
Maximum
[100%]
S
Use Set Point Maximum to establish, as a percentage of Full Scale Flow, the threshold
at which the output alarm is activated. Flow rates above the threshold activate the
outputalarm.
To change the maximum set point, follow these steps from the Inputs/Outputs menu:
1. Select Digital Output 1, 2, 3 or 4 and press [E] to open the Digital Output menu.
2. From the Digital Output menu select Set Point Maximum and press [E] to open
the Set Point Maximum display.
3. Enter the set point maximum value and press [E] to save the new parameter
and return to the Digital Output menu.
Using the Main Menu Programming Options
Page 42 March 2022MAG-UM-01272-EN-21