Communication
Diagnostics
Port A
Counters
[0]
S
Port counters are used for diagnostics when configured for Modbus RTU. These counters
are only cleared on power up.
Counter Description
Pkts Processed Number of packets processed by meter.
Broadcast Pkts Number of broadcast packets (address = 0) processed by meter.
CRC Errors Number of received packets with CRC error; packet is discarded.
Pkts Rcvd Number of packets received with an address of the configured
portaddress.
Pkts Sent Number of packets transmitted in response to a received packet.
Parity Errors Number of characters with parity errors (for example, if the received
character has a mismatch between the number of 1s and its parity bit);
packet is discarded.
Framing Errors Number of characters with framing errors (for example, stop bit is not
found – indicates that synchronization with the start bit has been lost
and the character is improperly framed); packet is discarded.
Overrun Errors Number of characters received that were not processed due to
degradation of system performance.
Break Detects Number of detections that transmission line is locked (for example, the
receive line is low for 10-bit transmissions following a missing stop bit).
Port B
Counters
[0]
S
Daughterbrd
config.
Parameter Value Description
BACnet MS/TP Baud 9600 Baud Rate of the BACnet MS/TP network. Common settings would be
9600, 19200, 38400, and 76800.
BACnet MS/TP MAC ID 1 Sets BACNET MS/TP Device ID of module/meter. Max value = 127.
BACnet Max Master 127 Sets the max master variable for the device. Max value = 127.
BACnet Instance 10001 Sets the BACnet Instance number. The instance number is an unsigned
decimal number that can range from 0 to 4,194,302. Every device on a
BACnet network gets an instance number, and two devices must not have
the same number.
Advanced
Data Logger
OTE: N This
feature needs an
additional memory
token that is not
included with the
standard meter
The Data Logger feature requires firmware version 1.10 or later. Reference BadgerMeter
P/N 67354-003 to obtain a firmware upgrade kit. See the M2000 Data Logging user manual, available
at www.badgermeter.com, for details on this feature.
The Data Logging feature records three types of events to a memory token:
• Totalizer/error events
• Configuration change events
• Startup events (power up, power down or reset events)
Token Copy
OTE: N This
feature needs an
additional memory
token that is not
included with the
standard meter
• Configuration
• Store to Token
• Restore to Token
See the M2000 Store/Restore user manual, available at
www.badgermeter.com, for details on using the Token Copy features.
Using the Main Menu Programming Options