Advanced
Analog Calibrate Custom Settings
[Zero Scale: 0 mA]
[Full Scale: 0 mA]
S
To set the analog calibration custom settings, follow these steps from the
Advanced menu:
1. Select Analog Calibrate to view the Analog Calibrate menu.
2. Select Custom Settings to view the Custom Settings display.
3. Select one of the following:
Offset 4 mA
Offset 20 mA
4. Configure the offset.
5. Press [E] to save the option and to return to the Custom
Settings menu.
6. Press [E] to return to the Analog Calibrate menu.
Factory Settings
[Factory Set]
S
To change the analog calibration factory settings, follow these steps from
the Advanced menu:
1. Select Analog Calibrate to view the Analog Calibrate menu.
2. Select Factory Settings to view the Factory Settings display.
3. Select one of the following:
Calibration Point A
Calibration Point B
4. Set the calibration point to the measured output current.
5. Press [E] to save the option and to return to the Factory
Settings menu.
6. Press [E] to return to the Analog Calibrate menu.
Software Filter
MDN-Filter
Description This software filter operates as a median filter. This filter is very responsive
and can be used to help stabilize flow measurements. It is enabled by
selecting a non-zero filter size. Supported filter sizes are:
S0 - Size 0
S5 - Size 5
S7 - Size 7
S9 - Size 9
The filter technique uses the median value of the last Sx samples used for
determining flow measurement.
Using the Main Menu Programming Options
Page 50 March 2022MAG-UM-01272-EN-21