Peak Detect offers a diagnostic view of the acceleration components observed
during flow conditions. This parameter records the “high water mark” of the
measured accelerations component. This value helps to properly configure
the Acceleration Factor parameter. Generally, you set the acceleration factor at
about 75% of the Peak Detect measurement.
To reset the Peak Detect setting, follow these steps from the Advanced menu:
1. Select Peak Detect from the Advanced menu.
2. Press [+] to reset.
3. Press [E] to return to the Advanced menu.
Software Filter
ZFS-Filter
Description This software filter operates as a zero-flow stability filter. A specific volume is
defined for a specific time window. If that volume is not measured during the
time window, this volume is ignored and not totalized. The actual flow can be
monitored with the status function. Filter options are:
• Volume USG
• Time
• Status (shows Volume and Time)
Software Filter
IIR-Filter
Description This software filter operates as an infinite impulse response filter, used to help
suppress erratic flow measurements. Contact BadgerMeter Technical Support.
Activation ON/OFF
Coefficient Min Numeric entry
Coefficient Max Numeric entry
Coefficient Status Numeric entry
Sensitivity Numeric entry
Hysteresis m/s