Empty Pipe Cal.
[Default]
S
Fluid conductivity impacts the performance of empty pipe measurements. If you require empty pipe
detection, you should perform this empty pipe calibration procedure.
Before starting the empty pipe calibration, verify that empty pipe detection is enabled. Also, run both
the empty pipe and the full pipe calibration procedures.
Calibrating an Empty Pipe
Before calibrating an empty pipe, verify that the pipe is empty.
To calibrate with an empty pipe, follow these steps from the Advanced menu:
1. Select Empty Pipe Cal to view the Calibration menus.
2. Select Cal. Empty Pipe to view the Empty Pipe Calibrate menu.
3. To enable calibration, place the cursor on the calibration enable line and press [E].
4. Wait 30 seconds for voltage measurement to stabilize.
5. To save the setting, place the cursor on Exit with Save and press [E].
Calibrating a Full Pipe
Before calibrating a full pipe, verify that the pipe is full.
To calibrate with a full pipe, follow these steps from the Advanced menu:
1. Select Empty Pipe Cal to view the Calibration menus.
2. Select Cal. Full Pipe to view the Full Pipe Calibrate menu.
3. Enable calibration by placing the cursor on the calibration enable line and press [E].
4. Wait 30 seconds for voltage measurement to stabilize.
5. To save the setting, place the cursor on Exit with Save and press [E].
Using the Main Menu Programming Options
Page 53 March 2022 MAG-UM-01272-EN-21