Ultrasonic washer
Model: E0201
During the security routine, a display with information on the
canceled cycle and the reason for the cancellation is changed.
The reason for canceling the cycle is also printed on the
printing belt.
The reason for the canceled cycle, the temperature and the
estimated time for the end of the routine if it changes in the
safety routine screen. An audible alarm will be activated and
you can deactivate it at any time by touching the "Deactivate
alarm" icon.
After performing the security routine, the cycle ends with the
message Canceled cycle due to a fall.
The fall will print on the printing belt.
The "Cycle time" indicates the total time and the "End of cycle
time" indicates the exact time in which the cycle was canceled.
At the end of the cycle, a bell will activate to inform the operator
about the end of the cycle. For more information, see the topic
"End of the cycle".
If there is a fall that precludes the start of a cycle, it will be
indicated in the same way in the calf and the cycle in the
beginning that until the fall is corrected.
13.1 Description of alarms
Alarms
Description
Over temperature alarm
This alarm occurs when the temperature exceeds the value set
at +10 ° C.
Low temperature alarm
This alarm occurs when the temperature is below 5 ° C of the
programmed value.
Temperature sensor alarm
This alarm occurs when there are errors in the temperature
sensor reading.
Supply alarm
This alarm is produced when the water level required for the
process is not reached.
Drain alarm
This alarm occurs when the bath does not drain.
Supervision time alarm
This alarm occurs when the heating reaches the value
programmed in a determined time.
Cycle with restoring parameters alarm
This alarm occurs when the parameters are restored that
prevent the cycle from starting.
Close the cover for cycle start
This alarm is produced when an attempt is made to start a cycle
with the cover open, if it opens during the cycle.
Ultrasound system failure
It occurs when the absence of ultrasound is noted during the
cycle.