Instructions for Use
Code: IU.01.09.0007
Version / Revision: 6.001
Page: 90 de 143
Ultrasonic washer
Model: E0201
First aid measures in the use of products used in Ultrasonic washers E0201
In the case of:
Contact with the skin: pollute the clothes and contaminated shoes. Wash the skin with plenty of water
for at least 20 minutes;
Contact with eyes: wash eyes with running water. Wear contact lenses if you wear them after the first
5 minutes and continue washing your eyes for at least 15 minutes. Consult an immediate doctor;
Ingestion: if swallowed, seek medical attention. It does not induce vomiting;
Inhalation: remove the victim from the air. In the case of signs of intoxication, seek medical
assistance.
Main symptoms and effects, acute or delayed: Contact can cause irritation in eyes or mucous membranes.
Notes to the doctor: symptomatic treatment.
Precaution
The possible risks are of the same nature as any electromechanical equipment, that is, electrical discharges and injuries
caused by accidents caused by the lack of attention on the part of the operator or the fall of the operation.
To carry out maintenance, turn off the power supply;
Never ultrasonic without liquids inside the tank;
Never place objects or objects to be cleaned in direct contact with the bottom of the tank, as it may
cause permanent damage to the tank.
Warnings
The time required for each process is variable, such as the number of pieces, the level of cleanliness
of the cleaning solution, the ambient temperature, among others, therefore, the time for each situation
must be analyzed by the user;
The operator must wear hearing protectors so that the noise in the ear does not cause future hearing
problems.
Possible adverse effects.
The lack of connection to the land, as well as the poor dimensioning of the electrical network could compromise the
safety of the operator.
Special attention and clarifications on the use of medical equipment.
At the end of the day, the E0201 Ultrasonic washer must be disconnected from all supply lines.
On the power supply panel, turn off the circuit breakers that power the water pump motor and control.
Close the water supply valve.
The information presented here does not replenish the recommendations presented by the
manufacturer of the product.
16. Problems and solutions
Problem
Cause
Corrective measure
Poor water supply.
• The water valves are closed.
• Open the valves located near the water inlet.
The water supply hoses are
doubled.
Realign hoses or replace them.
• Low water pressure.
• Install a pressurizer for 2 to 3 bars.