Ultrasonic washer
Model: E0201
• The power supply is not
connected.
• The general circuit breaker is off.
• Perform the electrical connection to the
equipment.
• Turn on the circuit breaker.
Emission of extraordinary
sounds, with the equipment
in operation.
• The utensils and instruments are
packed irregularly.
• The components of the equipment
are subtle and unbalanced.
• Abort the cycle and redistribute utensils
and instruments correctly.
• Check components such as: pumps,
transducers and valves.
The foam overflows from the
bath.
• Detergent used for specific
purposes.
• Incorrect amount of detergent.
• Abort the cycle. Flush the chamber to
eliminate excess foam and change
detergent.
• Abort the cycle. Flush the chamber to
eliminate excess foam and stabilize the ideal
detergent quantity.
• Selected program.
• Utensils and / or incorrectly
disposed instruments.
• Insufficient amount of detergent.
• Injector nozzles are blocked and /
or blocked.
• Understanding the selected program is
ideal for the material and its success.
• Redistribute utensils and instruments
correctly.
• Repeat the wash changing the amount of
detergent in the program.
• Clean the mouthpieces and, if necessary,
replace them.
The command screen goes
out.
• Control circuit breaker tripped.
• Connector between interface and
flow control.
• Adjust the circuit breaker.
• Free to fit the connector.
Opening and closing
movement is difficult.
• Lack of adjustment in the hinge.
• Carry out maintenance on the hinge.
The slap is closed, however,
the interface says it is open.
• Fall to adjust the micro cover.
• Carry out maintenance and correction in
the position of the micro box.
The printer does not print.
• Defective printer.
• Source of defective printer.
• Fall of communication with the
controller.
• Perform maintenance and replenishment of
components.
• Carry out maintenance or change the fuel.
• Check the connection with the connection
cables.
• Communication fall between the
monitor and the controller.
• Defective controller.
• Check the communication cables.
• Perform maintenance or replenish
components.
Water does not reach the
selected temperature.
• Resistance to burning.
• Fuse blown.
• Fall of the temperature sensor.
• Temperature sensor incorrectly
placed on the flow.
• Replenish resistance.
• Replace the fusible.
• Check the replacement of the temperature
sensor.
• Valve to place the sensors.
Temperature reading error.
• Flow temperature sensor.
• Defective temperature sensor.
• The controller cannot be connected.
• Free to read the sensor.
• Perform the replenishment of components.
• Check the connection to the controller.
Corrective measures must be carried out by an authorized technician. In case of non-resolution of the
problem or other problems, contact Medixx or its service channels.