7 Main Menu Device Setup LB 470 Level
56925BA2 Rev. 03, 11/2017
5. Click on the < Start > button (Fig. 96, item 5).
The measurement starts.
6. Click < Ok > to accept the value.
The level has been adjusted to the process value (see standard display).
When executing an Adjust, the measurement parameter set was overwritten.
If the system should be recalibrated, a Recall of the measurement parameter
set to the calibration parameter must be performed. Otherwise the Adjust will
become lost.