BD / BF / ED / FD / FED (E3.1) 02/2022 page 23/116
The domestic socket must also provide a protective conductor. Make sure that the connection of the
protective conductor of the domestic installations to the chambers protective conductor meets the latest
technology. The protective conductors of the socket and plug must be compatible!
DANGER
Electrical hazard due to missing protective conductor connection.
Deadly electric shock.
Make sure that the chamber’s power plug and the power socket match and se-
curely connect the electrical protective conductors of the chamber and the house
installation.
Only use original connection cables from BINDER according to the above specification.
UL chambers: Use only a UL Listed Power supply cord (UL category ELBZ) according to the above
specification. For outside USA use a certified power supply cord according to national requirements.
Prior to connection and start-up, check the power supply voltage. Compare the values to the specified
data located on the chamber’s type plate (left-hand side of the chamber, chap. 1.6).
NOTICE
Danger of incorrect power supply voltage due to improper connection.
Damage to the chamber.
Check the power supply voltage before connection and start-up.
Compare the power supply voltage with the data indicated on the type plate.
When connecting, please observe the regulations specified by the local electricity supply company as
well as the local or national electrical regulations (VDE directives for Germany).
Observe a sufficient current protection according to the number of devices that you want to operate. We
recommend the use of a residual current circuit breaker.
BF, FD, FED: Do not place the power cable over the door gap when the chamber is hot after operation.
Pollution degree (acc. to IEC 61010-1): 2
Over-voltage category (acc. to IEC 61010-1): II
See also electrical data (chap. 16.4 to 16.7).
To completely separate the chamber from the power supply, you must disconnect the power
plug. Install the chamber in a way that the power plug is easily accessible and can be easily
pulled in case of danger.