BD / BF / ED / FD / FED (E3.1) 02/2022 page 83/116
Different electrical data for FED-UL constructed for the USA and Canada
(model versions FED056UL-120V, FED115UL-120V, FED260UL-240V, FED720UL-208V)
voltage
at 50 Hz power frequency
V 120 120 240 208
at 60 Hz power frequency
V 120 120 240 208
Power plug of the power cable
Chamber fuse (external)
5x20 mm / 250V / time-lag (T)
A 12.5 12.5 --- ---
Miniature circuit breaker (combina-
tion element, external)
A --- --- 16 ---
Miniature circuit breaker (internal)
Environment-specific data
Energy consumption at 150 °C /
302 °F
300 340 420 800
All technical data is specified for unloaded chambers with standard equipment at an ambient temperature
of +22 °C +/- 3 °C / 71.6 °F +/- 5.4 °F and a power supply voltage fluctuation of +/-10%. Technical data is
determined in accordance to BINDER Factory Standard Part 2:2015 and DIN 12880:2007.
All indications are average values, typical for chambers produced in series. We reserve the right to
change technical specifications at any time.
If the chamber is fully loaded, the specified heating up times may vary according to the load.
16.9 Equipment and options (extract)
To operate the chamber, use only original BINDER accessories or accessories / components
from third-party suppliers authorized by BINDER. The user is responsible for any risk arising
from using unauthorized accessories.
Microprocessor temperature controller
One timer function: Delayed Off
Three timer functions: Delayed On, Delayed Off and
Temperature dependent Delayed Off
√ √
Temperature safety controller class 3.1 acc. to DIN
12880:2007
√ √
Temperature safety controller class 2 acc. to DIN 12880:2007
USB interface to read out the measured values
Communication interface Ethernet
Exhaust duct, internal diameter 50 mm /
1.97 inches
, with ad-
justable ventilation slide
√ √ √ √ √
Adjustable air change by means of rear exhaust duct (50
mm)
√ √ √ √ √