BD / BF / ED / FD / FED (E3.1) 02/2022 page 75/116
16.4 BD technical data
Height, gross (incl. feet)
Depth, gross (incl. door handle, ex-
haust duct)
mm / inch
640 / 25.20 680 / 26.77 815 / 32.09 870 / 34.25
Wall clearance rear (minimum)
Wall clearance side (minimum)
Exhaust duct, outer diameter
Number of inner glass doors
Quantity of racks (regular)
Max. load per rack (standard rack)
Max. load per rack
(optional perforated shelf)
Kg / lbs 35 / 77 35 / 77 35 / 77 35 / 77
Max. load per rack
(optional heavy load rack)
Kg / lbs 50 / 110 70 / 154 80 / 176 70 / 154
Temperature
range
from … degrees
above ambient
°C / °F 5 / 9 5 / 9 5 / 9 5 / 9
Temperature fluctuation
at 37 °C / 98.6 °F
± K 0.2 0.1 0.2 0.1
Temperature uniformity (variation)
at 37 °C / 98.6 °F
± K 0.4 0.4 0.4 0.7
Heating up time to 37 °C / 98.6 °F
Recovery time after door was
opened for 30 sec at 37 °C / 98.6 °F
minutes 16 16 19 23
Electrical data (model versions BD056-230V, BD115-230V, BD260-230V, BD720-230V)
System of protection acc. to EN
60529
IP 20 20 20 20
voltage
at 50 Hz power frequency V 230 230 230 230
at 60 Hz power frequency V 230 230 230 230
Nominal power with option internal
socket
kW 0.80 0.85 1.35 2.15