BD / BF / ED / FD / FED (E3.1) 02/2022 page 75/116
16.4 BD technical data
Chamber size
BD 56
BD 115
BD 260
BD 720
Exterior dimensions
Width, net
mm / inch
560 / 22.01
710 / 27.95
815 / 32.09
1165 / 45.87
Height, gross (incl. feet)
mm / inch
625 / 24.60
735 / 28.94
965 / 37.99
1590 / 62.60
Depth, net
mm / inch
565 / 22.24
605 / 23.82
760 / 29.92
816 / 32.13
Depth, gross (incl. door handle, ex-
haust duct)
mm / inch
640 / 25.20 680 / 26.77 815 / 32.09 870 / 34.25
Wall clearance rear (minimum)
mm / inch
160 / 6.30
160 / 6.30
160 / 6.30
160 / 6.30
Wall clearance side (minimum)
mm / inch
100 / 3.94
100 / 3.94
100 / 3.94
100 / 3.94
Exhaust duct, outer diameter
mm / inch
52 / 2.05
52 / 2.05
52 / 2.05
52 / 2.05
Doors
Number of doors
1
1
1
2
Number of inner glass doors
1
1
1
2
Interior dimensions
Width
mm / inch
360 / 14.17
510 / 20.08
610 / 24.02
960 / 37.80
Height
mm / inch
420 / 16.54
530 / 20.87
760 / 29.92
1280 / 50.39
Depth
mm / inch
380 / 14.96
420 / 16.53
550 / 21.65
605 / 23.81
Interior volume
l / cu.ft.
57 / 2.01
112 / 3.96
255 / 9.01
737 / 26.03
Steam space volume
l / cu.ft.
63 / 2.22
127 / 4.49
279 / 9.85
791 / 27.93
Racks
Quantity of racks (regular)
2
2
2
2
Quantity of racks (max.)
4
5
8
16
Max. load per rack (standard rack)
Kg / lbs
30 / 66
30 / 66
40 / 88
45 / 99
Max. load per rack
(optional perforated shelf)
Kg / lbs 35 / 77 35 / 77 35 / 77 35 / 77
Max. load per rack
(optional heavy load rack)
Kg / lbs 50 / 110 70 / 154 80 / 176 70 / 154
Permitted total load
Kg / lbs
70 / 154
150 / 330
270 / 595
315 / 694
Weight
Weight (empty)
Kg / lbs
38 / 84
54 / 119
85 / 187
170 / 375
Temperature data
Temperature
range
from … degrees
above ambient
°C / °F 5 / 9 5 / 9 5 / 9 5 / 9
up to
°C / °F
100 / 212
100 / 212
100 / 212
100 / 212
Temperature fluctuation
at 37 °C / 98.6 °F
± K 0.2 0.1 0.2 0.1
Temperature uniformity (variation)
at 37 °C / 98.6 °F
± K 0.4 0.4 0.4 0.7
Heating up time to 37 °C / 98.6 °F
minutes
45
55
65
70
Recovery time after door was
opened for 30 sec at 37 °C / 98.6 °F
minutes 16 16 19 23
Electrical data (model versions BD056-230V, BD115-230V, BD260-230V, BD720-230V)
System of protection acc. to EN
60529
IP 20 20 20 20
Nominal
voltage
(+/-10%)
at 50 Hz power frequency V 230 230 230 230
at 60 Hz power frequency V 230 230 230 230
Current type
1N~
1N~
1N~
1N~
Nominal power
kW
0.30
0.35
0.85
1.65
Nominal power with option internal
socket
kW 0.80 0.85 1.35 2.15