EasyManua.ls Logo

Binder KB - Setting the Safety Controller

Binder KB
147 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
KB / KB-UL (E4 + E6) 10/2021 page 105/147
Fixed value
08.03.2015 05:05:06
Submenu “Safety controller
(view with optional chamber equipment)
Select “Mode”
and press the operating button.
..\ Safety controller
Mode
Overtemperature
Undertemperature
Show settings
Close Home
Fixed value
08.03.2015 05:05:06
Submenu “Safety controller mode”.
The current safety controller mode is displayed:
“Limit (absolute)” or “Offset (relative)”
To change the mode, press the operating button.
The modified safety controller mode is displayed.
..\ Safety controller mode
Limit (absolute)
Close Home
When changing the safety controller mode, the safety controller setpoint which had been ac-
tive before in this mode becomes active again.
Entering the safety controller setpoint
Fixed value
08.03.2015 05:05:06
Submenu “Safety controller
(view with standard chamber equipment)
Select “Setpoint”
and press the operating button.
..\ Safety controller
Mode
Setpoint
Show settings
Close Home