KB / KB-UL (E4 + E6) 10/2021 page 107/147
The overview display shows the set-points and actual values of the main temperature controller and the
safety controller and indicates the safety controller mode.
08.03.2015 05:05:06
Overview display with safety controller mode “Limit”
(example values).
In case of a temperature safety device class 3.1
combined with optional chamber equipment, “Safety
controller” is indicated as “Overtemperature” in-
stead.
Chamber temperature [°C]
37.0 37.6
Safety controller mode
Limit (absolute)
Safety controller [°C]
Home
08.03.2015 05:05:06
Overview display with safety controller mode “Off-
set” (example values).
In case of a temperature safety device class 3.1
combined with optional chamber equipment, “Safety
controller” is indicated as “Overtemperature” in-
stead.
Chamber temperature [°C]
Safety controller mode
Offset (relative)
Safety controller [°C]
Home
Go back to the initial view with “Home”.