Symbol on the type plate
Information
(KT-UL only)
The equipment is certified by Underwriters Laboratories Inc.
®
according to standards CAN/CSA-C22.2 No. 61010-1, 2
nd
Edition, 2004-07 (Electrical Equipment for Measurement,
Control, and Laboratory Use; Part 1: General Requirements); UL
61010-1, 2
nd
Edition, 2005-07-22 (Electrical Equipment for
Measurement, Control, and Laboratory Use; Part 1: General
Requirements); IEC 61010-1:2010, 3
rd
Edition and IEC 61010-2-
10:2003 (Particular Requirements for Laboratory Equipment for
the heating of materials).
1.5 General safety instructions on installing and operating the cooling
incubator
With regard to operating the cooling incubator KT and to the installation location, please observe the
guideline BGI/GUV-I 850-0 on safe working in laboratories (formerly BGR/GUV-R 120 or ZH 1/119
laboratory guidelines issued by the employers’ liability insurance association) (for Germany).
BINDER GmbH is only responsible for the safety features of the unit provided skilled electricians or
qualified personnel authorized by BINDER perform all maintenance and repair, and if components
relating to chamber safety are replaced in the event of failure with original spare parts.
To operate the unit, use only original BINDER accessories or accessories from third-party suppliers
authorized by BINDER. The user is responsible for any risk caused by using unauthorized accessories.
CAUTION
Danger of overheating.
Damage to the unit.
Do NOT install the unit in unventilated recesses.
Ensure sufficient ventilation for dispersal of the heat.
Do not operate the cooling incubator KT in hazardous locations.
DANGER
Explosion hazard.
Danger of death.
Do NOT operate the unit in potentially explosive areas.
KEEP explosive dust or air-solvent mixtures AWAY from the unit.
The cooling incubator KT does not dispose of any measures of explosion protection.
DANGER
Explosion hazard.
Danger of death.
Do NOT introduce any substance into the cooling incubator which is combustible or
explosive at working temperature.
NO explosive dust or air-solvent mixture in the inner chamber.

Table of Contents