KT (E6.1) 07/2015 Page 125/138
The technical data refers to the defined usable volume.
Do NOT place samples outside this usable volume.
Do NOT load this volume by more than half to enable sufficient airflow inside the chamber.
Do NOT divide the usable volume into separate parts with large area samples.
Do NOT place samples too close to each other in order to permit circulation between them
and thus obtain a homogenous distribution of temperature and humidity.
23.4 KT (E6.1) technical data
Depth (incl. door handle, instrument triangle,
connection)
mm / inch 695 / 27.36 720 / 28.35 720 / 28.35
Wall clearance rear (minimum) (spacer)
Wall clearance side (minimum)
Quantity of inner glass doors
Quantity of racks standard / max.
4 °C / 23 °F up to +100 °C / 212 °F
Temperature uniformity
(variation)
Max. heat compensation at 40 °C / 104°F
Recovery time after doors
were open for 30 sec.
Electrical data
(model versions KT053-230V, KT115-230V, KT240-230V)
IP System of protection acc. to EN 60529
Nominal voltage
(+/-10%)
Energy consumption
Overvoltage category acc. to IEC 61010-1
Pollution degree acc. to IEC 61010-1
Unit fuse 5x20 mm 230V / middle-time-lag M