19.8 Water protected internal socket (option)
You can turn on and off the interior socket via the controller menu (chap. 7.2).
The internal socket is splash proof.
IP system of protection 67, 100-240 V 1N ~ 50-60 Hz
Charge max. 500 W
Maximum permitted operating temperature with this option: 90 °C / 194 °F.
WARNING
Exceeding the permitted maximum temperature.
Electrical hazard.
Danger of death.
Damage to the internal socket.
∅ Do NOT exceed the temperature set-point of 90 °C / 194 °F.
Set the overtemperature safety controller to “Limit” and a maximum value of
90 °C / 194 °F.
Heat emission of electrical devices connected inside the chamber may modify the temperature
range.
Risk of short circuit.
Damage to the unit.
Use the supplied plug only (IP protection type 67). Plug it in and
contact.
If the socket is not used, close the screw cap and turn it to secure.
Turning on and off the voltage of the interior socket is possible at the chamber controller through Menu >
Optional equipment > Interior socket, see chap. 7.2. Turning off the incubator at the main power switch
also switches the interior socket voltage-free.
Figure 12: Innenraumsteckdose
(front view)