82
Note: Insure pump is properly secured before attaching fluid bag to short stand
and connecting to patient to assure pump cannot fall and harm patient.
Battery-Pack SP with Wireless (Li-Ion) (8713182U)
For further information on the Battery Pack with wireless (Li-Ion), see chapter
6.3 "Battery Pack SP with WiFi".
Battery-Pack SP without Wireless (NiMH) (8713180A)
For further information on the Battery Pack without wireless (NiMH), see chap-
ter 6.1 "Battery Maintenance".
Interface Lead CAN SP (8713230)
Interface Lead CAN SP is needed in order to set up a connection between the
SpaceStation and /or pump and the computer to use the service program
(HiBaSed).
1.) Push plug into socket F3 on the SpaceStation or P2 on the pump and
connect with the CAN/USB converter.
2.) Connect CAN/USB converter to comp
ute
r USB port as described in the
HiBaSed Instructions for Use manual.
Caution: The Interface Lead CAN SP is only to be used by the service
department; never use while patient is connected.
Note: A maximum of three plugs can be stacked upon each other in socket P2.
Connection Lead for Staff Call SP (8713232)
To connect the Infusomat® Space to a staff call system, use the Connection
Lead for Staff Call S
P.
Note: Test staff call signalling each time new pump connection is established.
The Infusomat® Space offers three different staff call operating modes. They are
displayed in the signalling scheme below. Consider the functioning of the staff
call in the hospital when choosing an operating mode. Choose the operating
mode via the service program.
Chapter 10
OPTIONAL SPACE ACCESSORIES