Brookeld Engineering Labs., Inc. Page 19 Manual No. M13-167
II.5.3 View Results
Testresults(datales)canbesavedtotheinternalmemoryoftheDV2TortoaUSBFlashDrive.
TheseslescanbereloadedintotheDV2Tforreview,analysis,orprintingthroughtheViewResults
function.AleofTestResultsthatissavedontoaUSBFlashDrivecanbeviewedonanyDV2T
Viscometer.
WithintheViewResultsfunction,theusercanaccesstheinternalmemoryoftheDV2TViscometer
oranyUSBFlashDrivethatisconnectedtoaUSBport .Theviscometerwillpointtothe
USB FlashDriveaccordingtotheorderinwhichitisconnected.TherstUSBFlashDrivethat
is connected will bereferredtoas#1onboththeViewResultsscreenandtheStatusBar.Youcan
haveasmanyasthreeUSBFlashDrivesconnectedtotheDV2Tatanytime.
Resultslesthataredisplayedonthescreencanbesortedbydateofcreationorbyalpha/numeric.
ThissortingisselectedbypressingtheNavigationIcon or .
TIP: YoucanusetheManageFilesfunction tomoveResultslesfrominternalmemoryto
aUSBFlashDrive.
II.5.4 Manage Files
ResultFilesandTestFilescanbemanagedintheinternalmemoryoronUSBFlashDrivesfromthe
ManageFilesfunction.Folderstructurescanbeaddedorchangedtoassistwithdatamanagement.
Filesmaybecopied,moved,renamedordeleted.AccesstothisfunctioncanbelimitedwhenUser
IDandLogInfunctionsareimplemented(seeSectionIV.4.2).
Filesthataredisplayedonthescreencanbesortedbydateofcreationorbyalpha/numeric.This
sortingisselectedbypressingtheNavigationIcon or .
II.5.5 External Mode
TheDV2TViscometercanbecontrolledfromacomputerthroughtheuseofoptionalBrookeld
SoftwareRheocalcT.TheviscometermustbeplacedintoexternalcontrolmodefromtheMain
Menu
.TheviscometermustbeconnectedtothecomputerwithaUSBAcable(DVP-202).
TIP: TheStatusBarwillindicateaproperconnectiontothecomputerbydisplayingtheComputer
Icon .
TheDV2TwilldisplayExternalModewhenconguredforoperationwiththecomputer.This
displayincludesaReturnbuttonthatwillresettheviscometertostandaloneoperation.
II.6 Range
The DV2T Viscometer will calculate the measurement range for a specic spindle and speed
combination.Thisinformationisdisplayedonthescreenwhileselectingthespindlenumberas
showninFigureII-6.TheRangeisalsoshownintheRunningViscosityTestviewduringthe
measurement.ViscositywillbedisplayedintheunitofmeasurespeciedinSettings and is
settocentipoise(cP)fromthefactory.