Brookeld Engineering Labs., Inc. Page 36 Manual No. M13-167
Eachdatapointincludes:Viscosity,Torque,Speed,Temperature,Time,ShearStress
(SS),ShearRate(SR),Density,andAccuracy.
Post Test Averaging : Average and Standard Deviation are calculated for measured and calculated
parametersincluding:Viscosity,Torque,ShearStress,andTemperature.
PostTestAveragingiscalculatedregardlessoftheDataCollectionsetting.
IfDataCollectionwassettoMultipointAveraging,thenthePostTest
Averagingwillcalculateanaverageoftheaverageddata.
WhenusingmultiplestepprogramsgeneratedthroughPGFlashsoftware,
theteststeputilizedforPostTestAveragingisspeciedintheReport
setting(seeSectionIII.7).
Test Used:Displaythetestelementsusedtogeneratethedataset.Inthisview,theCongure
Testbuttonisavailable.SelectingCongureTestwillprogramtheDV2Ttorun
thesametestutilizedtocollectthedatasetbeingviewedandpresenttheCongure
ViscosityTestscreen.
Notes:Documentanyrelevantinformationaboutthetestordata.Thisinformationwillbe
storedwiththedatasetoncesaved.
Compare Results: Load two data sets and view side by side.
Figure III-12
Device Info:DisplaybasicinformationaboutthedataleandthespecicDV2TViscometer
usedtocollectthedataincluding:dateandtimetheteststarted,completedand
wassaved;ifuseraccountsareactive,whichusersavedthedata;viscometerserial
number,rmwareversion,viscometerTorquerange(LV,RV,HA,HB,other).
III.10 Data Averaging
The DV2T Viscometer offers two techniques to average data, Live Averaging and Post Test
Averaging.Dataaveragingcanbeusefulwhenmeasuringsampleswithentrainedairorsuspended
particlesthatcasesomevariationinmeasurementresults.Dataaveragingmaynotbeusefulwhen
changesinmeasurementresultsarecausedbytherheologicalpropertiesofthetestsamplesuch