Brookeld Engineering Labs., Inc. Page 79 Manual No. M13-167
Appendix F - The Brookeld Guardleg
Theguardlegwasoriginallydesignedtoprotectthespindleduringuse.Therstapplicationsofthe
BrookeldViscometerincludedhandheldoperationwhilemeasuringuidsina55-gallondrum.Itis
clearthatunderthoseconditionsthepotentialfordamagetothespindlewasgreat.Originalconstruction
includedasleevethatprotectedthespindlefromsideimpact.EarlyRVguardlegsattachedtothedial
housingandLVguardlegsattachedtothebottomofthepivotcupwithatwistandlockmechanism.
ThecurrentguardlegisabandofmetalintheshapeoftheletterUwithabracketatthetopthatattaches
tothepivotcupofaBrookeldViscometer/Rheometer.Becauseitmustattachtothepivotcup,the
guardlegcannotbeusedwithaCone/Plateinstrument.AguardlegissuppliedwithallLVandRVseries
instruments,butnotwiththeHAorHBseries.It’sshape(showninFigureF-1andF-2)isdesignedto
accommodatethespindlesoftheappropriatespindleset;therefore,theRVguardlegiswiderthanthe
LVduetothelargediameteroftheRV#2spindle.Theyarenotinterchangeable.
ThecalibrationoftheBrookeldViscometer/Rheometerisdeterminedusinga600mLLowFormGrifn
Beaker.ThecalibrationofLVandRVseriesinstrumentsincludestheguardleg.Thebeakerwall(for
HA/HBinstruments)ortheguardleg(forLV/RVinstruments)denewhatiscalledthe“outerboundary”
ofthemeasurement.ThespindlefactorsfortheLV,RV,andHA/HBspindlesweredevelopedwiththe
aboveboundaryconditions.Thespindlefactorsareusedtoconverttheinstrumenttorque(expressed
asthedialreadingor%Torquevalue)intocentipoise.Theoretically,ifmeasurementsaremadewith
differentboundaryconditions,e.g.,withouttheguardlegorinacontainerotherthan600mLbeaker,
thenthespindlefactorsfoundontheFactorFindercannotbeusedtoaccuratelycalculateanabsolute
viscosity.Changingtheboundaryconditionsdoesnotchangetheviscosityoftheuid,butitdoes
changehowtheinstrumenttorqueisconvertedtocentipoise.Withoutchangingthespindlefactorto
suitthenewboundaryconditions,thecalculationfrominstrumenttorquetoviscositywillbeincorrect.
Practicallyspeaking,theguardleghasthegreatesteffectwhenusedwiththe#1spindlesofthe
LVandRVspindlesets(Note:RV/HA/HB#1spindleisnotincludedinstandardspindleset).Any
otherLV(#3)orRV(#3-#7)spindlecanbeusedina600mLbeakerwithorwithouttheguard
legtoproducecorrectresults.TheHAandHBseriesViscometers/Rheometersarenotsuppliedwith
guardlegsinordertoreducethepotentialproblemswhenmeasuringhighviscositymaterials.HA/
HBspindles#3through#7areidenticaltothosespindlenumbersintheRVspindleset.TheHA/HB
#1haveslightlydifferentdimensionsthanthecorrespondingRVspindles.Thisdimensional
differenceallowsthefactorsbetweentheRVandHA/HB#1spindlestofollowthesameratios
astheinstrumenttorqueeventhoughtheboundaryconditionsaredifferent.
Therecommendedproceduresofusinga600mLbeakerandtheguardlegaredifcultforsomecustomers
tofollow.Theguardlegisonemoreitemtoclean.Insomeapplicationsthe500mLoftestuid
requiredtoimmersethespindlesina600mLbeakerisnotavailable.Inpractice,asmallervesselmay
beusedandtheguardlegisremoved.TheBrookeldViscometer/Rheometerwillproduceanaccurate
andrepeatabletorquereadingunderanymeasurementcircumstance.However,theconversionofthis
torquereadingtocentipoisewillonlybecorrectifthefactorusedwasdevelopedforthosespecic
conditions.Brookeldhasoutlinedamethodforre-calibratingaBrookeldViscometer/Rheometer
toanymeasurementcircumstanceinMoreSolutionstoStickyProblems.Itisimportanttonotethat
formanyviscometerusersthetrueviscosityisnotasimportantasarepeatabledaytodayvalue.This
repeatablevaluecanbeobtainedwithoutanyspecialeffortforanymeasurementcircumstance.But,it