Brookeld Engineering Labs., Inc. Page 72 Manual No. M13-167
Appendix E - Calibration Procedures
TheaccuracyoftheDV2TisveriedusingviscositystandarduidswhichareavailablefromBrookeld
Engineering Laboratories or your local Brookeld agent. Viscosity standards are Newtonian, and
therefore, have the sameviscosity regardless of spindle speed (or shear rate). Viscosity standards,
calibrated at 25°C,areshowninTable E-1(SiliconeOils)andTable E-2 (MineralOils).
Formorehelpyoucangotothewebsite,
www.brookeldengineering.com,anddownloadthevideo.
Container size: ForViscosityStandards<30,000cP,usea600mLLowFormGrifnBeaker
havingaworkingvolumeof500mL.
ForViscosityStandards30,000cP,usetheuidcontainer.
InsideDiameter: 3.25”(8.25cm)
Height: 4.75”(12.1cm)
Note:Containermaybelarger,butmaynotbesmaller.
Temperature: Asstatedontheuidstandardlabel:(+/-)0.1°C
Conditions: TheDV2Tshouldbesetaccordingtotheoperatinginstructions.Thewaterbath
mustbestabilizedattesttemperature.ViscometerswiththelettersLV”or
RV”inthemodeldesignationmusthavetheguardlegattached(seeAppendix
Fformoreinformationontheguardleg).
Normal 25ºC Standard Fluids
Viscosity (cP) Viscosity (cP)
High Temperature Standard Fluids
Three Viscosity/Temperatures**
5 5,000 HT-30,000
10 12,500 HT-60,000
50 30,000 HT-100,000
100 60,000
500 100,000 **25ºC, 93.3ºC, 149ºC
1,000 Refer to Brookeld catalog for more information
Table E-1
MINERAL OIL VISCOSITY STANDARD FLUIDS
BEL Part No. Viscosity (cP) 25ºC
B29 29
B200 200
B600 600
B1060 1,060
B2000 2,000
B10200 10,200
B21000 21,000
B73000 73,000
B200000 200,000
B360000 360,000
Table E-2

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Brookfield DV2T and is the answer not in the manual?

Brookfield DV2T Specifications

General IconGeneral
TypeViscometer
ModelDV2T
Repeatability±0.2%
Data OutputUSB, RS-232
Accuracy±1.0% of range
Temperature RangeOptional temperature probe
Power Supply100-240V, 50/60Hz
Display5-inch Color Touch Screen
Viscosity Range1-2, 000, 000 cP

Related product manuals