EasyManua.ls Logo

Brookfield DV2T - III.8 Running A Test

Brookfield DV2T
91 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Brookeld Engineering Labs., Inc. Page 32 Manual No. M13-167
III.8 Running a Test
AviscositytestisstartedbypressingtheRunbuttonontheCongureViscosityTestscreen.When
Runispressed,thedisplaywillchangetotheRunningViscosityTestscreen(seeFigureIII-9).
Trend Bar
Command Keys
Status Bar
Title Bar
Test Name
Measurement Data
Figure III-9
TheRunningViscosityTestscreenprovidesinformationonthecurrentmeasurementincluding:
Torque,Viscosity,ShearStress,ShearRate,TemperatureandSpeed.
TorqueisthedeectionoftheViscometertorquesensor.Itisdescribedasapercent(%)andhas
arangeof0–100%.TheDV2Twillprovidemeasurementresultswithinthestatedaccuracy
providedtheTorquereadingisbetween10and100%.IftheTorquereadingfallsbelow10%the
labelsinthedataeldswillashtoindicateanerrorcondition.Brookelddoesnotrecommend
thatdatabecollectedbelow10%Torque;however,datacollectionisnotrestricted.
TIP: TorqueontheDV2TisequivalenttothedialreadingfromtheBrookeldDialReading
Viscometerorthe % readingfromBrookeldDigitalViscometers(DV-E, DV-I,
DV-II,DV-III).
ViscosityiscalculatedfromthemeasuredTorquebasedontheselectedspindleandspeedof
rotation.TheunitsofviscosityaredenedintheGlobalSettingssectionoftheSettingsMenu
.IftheTorquereadingfallsbelow10%,thelabelsinthedataeldswillashtoindicate
an error condition.
Shear StressiscalculatedfromthemeasuredTorquebasedontheselectedspindle.Theunitsof
shearstressaredenedintheGlobalSettingssectionoftheSettingsMenu
.IftheTorque
readingfallsbelow10%,thelabelsinthedataeldswillashtoindicateanerrorcondition.
TIP:ShearStresswillbedisplayedaszeroforspindlesthatdonothaveSRCvalues.
Shear Rateiscalculatedfromtheselectedspeedbasedontheselectedspindle.
TIP:ShearRatewillbedisplayedaszeroforspindlesthatdonothaveSRCvalues.

Table of Contents