Installation and Operation Manual
X-VA-MT3809G-MT3810G-eng
Part Number: 541B182AAG
July, 2018
Models MT3809G & MT3810G
2-5-1 How to Remove and Reinstall GP Housing Indicator Covers
Directions to remove and reinstall the GP Housing
Cover and adjust the Pointer:
1. To begin, make sure the oat is at rest and there isn’t
ow going through the meter.
2. Using a T20 TORX tool loosen the 4 cover screws.
Note, the screws will stay attached to the cover.
3. Remove the cover from the housing. The gasket should
still be attached to the cover in the gasket groove. If not,
place the gasket into the groove of the cover.
4. Using a at blade screwdriver with a 1/8" blade, hold the
pointer and turn the screw to align with the “R” on the scale.
It may take a few adjustments to get the pointer on the “R”.
5. To replace the cover, place the cover against the housing
and tighten the 4 screws using a T20 TORX tool. Tighten to
45 in-lbs or 5.08 N-m to keep a watertight seal.