Electronics
configuration
Function / signal Ui,V Ii, mA Pi, mW Ci, nF Li, ȝH Recommended Barrier #
Signal 4-20mA
(J1 terminals 12+ and 13-)
Pulse output
(J1 terminals 7+ and 8-)
Alarm circuits A Pepperl & Fuchs: KFA5-SR2-EX2.W
(J1 terminals 1+ and 2-)
KFA6-SR2-
EX2.W
10,5 13 34 §0 §0 Pepperl & Fuchs: KFD2-SR2-EX2.W
Alarm circuits B Pepperl & Fuchs: KFA5-SR2-EX2.W
(J1 terminals 4+ and 5-)
KFA6-SR2-
EX2.W
10,5 13 34 §0 §0 Pepperl & Fuchs: KFD2-SR2-EX2.W
Uo,V Iɨ, mA Po, mW Cɨ, ȝF Lo, mH Notes
Remote zero loop signal
(J1 terminals 10+ and 11-)
28 2,83 80 0.083 44
Ui,V Ii, mA Pi, mW Ci, nF Li, mH Recommended Barrier #
FOUNDATION Fieldbus loop
(J1 terminals 10+/11+ and 12-/13-)
Pulse output Pepperl & Fuchs: KFA5-SR2-EX2.W
(J1 terminals 5+ and 6-)
KFA6-SR2-
EX2.W
10,5 13 34 §0 §0 Pepperl & Fuchs: KFD2-SR2-EX2.W
Alarm circuits A Pepperl & Fuchs: KFA5-SR2-EX2.W
(J1 terminals 1+ and 2-)
KFA6-SR2-
EX2.W
Alarm circuits B Pepperl & Fuchs: KFA5-SR2-EX2.W
(J1 terminals 3+ and 4-)
KFA6-SR2-
EX2.W
Uo,V Iɨ, mA Po, mW Co uF Lo mH Notes
Remote zero loop signal
(J1 terminals 8+ and 9-)
8,03 0,81 6,5 8,4 1215
Ui,V Ii, mA Pi, mW Ci, nF Li, ȝH Recommended Barrier #
Inductive High Alarm circuits
(terminals «+» and «-»)
– for connection of circuits
Pepperl+Fuchs mod. SJ 3,5-SN
type 2
10,6 19,1 51 30 100
Pepperl & Fuchs:KFA5-SR2-EX2.W or
KFA6-SR2-EX2.W
Inductive Low Alarm circuits
(terminals «+» and «-»)
– for connection of circuits
Pepperl+Fuchs mod. SJ 3,5-SN
type 2
10,6 19,1 51 30 100
Pepperl & Fuchs:KFA5-SR2-EX2.W or
KFA6-SR2-EX2.W
4-20mA / HARTFoundation FieldbusInductive Alarms
§0
10,6 19,1 51 §0 §0
0
10,6 19,1 51 §0
10,6 19,1
10,6 19,1 51 §0 §0
FISCO barrier24 380 5320 0
51 §0 §0
10,6 19,1 51 §0 §0
Stahl Type : 9002/77-280-094-001
28 75 525 2,2 0.365 Stahl Type : 9001/01-280-075-101
28 84 660 §0 §0