Utility Requirements
62 / 120 H157654_9_011
System and Am-
plifiers
Mains Supply Power Con-
sumption
(kVA)*
Length of
Mains Cable
(m)
NEO OneBay 230V / 16A single phase 50 Hz,
or 208V 60 Hz / 20A single
phase in USA
1.3 5.5
NEO OneBay with
Heightening
230V / 16A single phase 50 Hz,
or 208V 60 Hz / 20A single
phase in USA
2.0 5.5
NEO TwoBay opti-
mized for solids
230V / 32A single phase 50 Hz
or 400V / 16A triple phase 60
Hz, or 208V 60 Hz / 30A single
phase in USA
3.0 5.5
NEO TwoBay opti-
mized for Mi-
croImaging Sys-
tems
230V / 32A single phase 50 Hz
or 400V / 16A triple phase 60
Hz, or 208V 60 Hz / 30A single
phase in USA
3.3 5.5
NEO NanoBay 230V / 16A single phase 50 Hz,
or 208V 60 Hz / 20A single
phase in USA
1.0 5.5
NMR Computer HP
and two LCD Dis-
plays
230V / 10A single phase 50 Hz,
or 208V 60 Hz / 20A single
phase in USA
0.13 ---
Bayvoltex Chiller for
MicroImaging Sys-
tems
230V 50/60 Hz / 16A single
phase or 208V 60 Hz / 20A sin-
gle phase in USA
~0.45 ---
BCU-I 230V 50/60 Hz / 16A single
phase or 208V 60 Hz / 20A sin-
gle phase in USA.
0.38 ---
BCU-II 230V 50/60 Hz / 16A single
phase or 208V 60 Hz / 20A sin-
gle phase in USA (power must
come from separate outlet).
0.62 ---
BCU 20 230V 50/60 Hz / 16A single
phase or 208V 60 Hz / 20A sin-
gle phase in USA
1.15
(with heating)
0.27
(with cooling)
---
UPS for (CCU) UPS requirements: At least 5 kW for the AVANCE NEO and CCU.
UPS for AVANCE
NEO Spectrometer
Cabinet
UPS requirements: At least 2.6 kW for the spectrometer cabinet
(depends on configuration). The battery time must be selected ac-
cording to the maximum duration anticipated for a power failure.
Note: For the electrical specifications associated to the CryoProbe, CryoProbe - Prodigy,
and the Bruker Nitrogen Liquifier accessories, refer to the relevant pages in CryoProbe
and Other Accessories [}91].
Table8.1: Requirements of Basic System (2 Channels)