CryoProbe and Other Accessories
96 / 120 H157654_9_011
Ambient operating temperature 5–28 ºC.
Figure10.4: Helium Compressor - Indoor Water Cooled
10.2.1.2 Helium Compressor - Indoor Air Cooled
The room air handling system must be able to dissipate 7.5 kW of heat.
Siting this helium compressor in the same room as the NMR is not recommended.
Ambient operating temperature 5–28 ºC.
Figure10.5: Helium Compressor - Indoor Air Cooled