CryoProbe and Other Accessories
H157654_9_011 99 / 120
Sumitomo Type: F-70H (Europe) F-70L (USA) CNA-61D
Power Consumption
Cool Down max. 50/60
Hz:
Steady State max.
50/60 Hz:
<8.0/<9.0 kW
6.9-7.2/7.9-8.2 kW
8.0/9.2 kW
7.5/8.5 kW
Acoustic Noise: Maximum < 75 dB(A) at 1m, outdoor unit only.
Ambient Operating
Temperature:
-30 to 45°C (-22 to
112°F)
-30 to 45°C (-22 to
112°F)
-30 to 45°C (-22
to 112°F)
Table10.4: Technical Data for Outdoor Compressors
10.3 Helium Cylinders
The next step in the site planning for a CryoProbe accessory involves determining the loca-
tion for the helium gas cylinder. A research grade helium cylinder (grade 6.0, 99.9999%) is
supplied by the customer and is connected to the CryoCooling unit. The cylinder must be out-
side the 5 Gauss line in a serviceable location (i.e. changing the bottle should not interfere
with nearby magnets). The helium regulator and a special He gas hose are supplied by
Bruker.
Two lengths for the He gas hose line are available (10 m and 20 m).
The helium compressor and helium gas cylinder may be sited in a remote room or in the
same room as the spectrometer as shown in the layout examples.
10.4 Summary of CryoProbe Options
Helium Compressors (3 Options)
• Indoor Air Cooled
• Indoor Water Cooled
• Outdoor Air Cooled
Helium Gas Transfer Lines
For Indoor Helium Compressors (3 Options)
• 6 m
• 10 m
• 20 m