CryoProbe and Other Accessories
100 / 120 H157654_9_011
For Outdoor Helium Compressors (11 Options)
• 3 m indoor line / 10 m outdoor line
• 3 m indoor line / 20 m outdoor line
• 3 m indoor line / 30 m outdoor line
• 6 m indoor line / 10 m outdoor line
• 6 m indoor line / 20 m outdoor line
• 6 m indoor line / 30 m outdoor line
• 10 m indoor line / 10 m outdoor line
• 10 m indoor line / 20 m outdoor line
• 10 m indoor line / 30 m outdoor line
• 20 m indoor line / 10 m outdoor line
• 20 m indoor line / 20 m outdoor line
Helium Cylinder High Pressure Gas Line (2 Options)
• 10 m
• 20 m
Helium Transfer line from CryoCooling Unit to CryoProbe
• A standard length transfer line, determined by the magnet, is delivered with each Cry-
oPlatform. Longer transfer lines may be ordered if necessary to accommodate the Cry-
oCooling unit in your laboratory. If a longer transfer line is required a special review of
your site will be required.
CryoPlatform
CryoCooling Unit
• Located outside 50 Gauss (5 mT) line.
• Must be serviceable from all sides.
• Must be at least 0.5 meters from walls on all sides.
Helium Compressor- Indoor Water or Air Cooled
• Located outside 5 Gauss line.
• Minimum room space is 1.25 m (width) x 0.8 m (depth) x 0.7 m (height).
• Must be serviceable from top and left side.
• The distance between the CryoCooling unit and the helium compressor component is de-
fined by the helium transfer lines ordered.
• Chilled water lines must be supplied to helium compressor (water cooled unit only).
• Ambient operating temperature 5º–28ºC.
Note:
The indoor helium compressor has rollers and is connected to flexible helium transfer lines.
Water-cooled models are additionally connected to water lines supplied by the customer
(typically high pressure flexible water tubing).