EasyManua.ls Logo

Bruker BPHG 90 - Page 21

Bruker BPHG 90
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Safety
Z33081_00_001 21
DANGER
DangerofSuffocationfromHydrogenGas
If excessive amounts of H2 gas is released in an enclosed space, hydrogen will
decreasetheamountofavailableoxygenandmaycausesuffocation.
1. The device should be installed in a sufficiently spacious, well ventilated room
equippedwithapermanentH2alarmgassensor.
2. The user is referred to MSDS safety data sheets for hydrogen gas for further
information.
WARNING
Dangerofinjuryfromelectricalshock!
Alifethreateningshockmayresultwhenthehousingisopenduringoperation.
1. Onlyqualifiedpersonnelshouldopenthehousing.
2. Disconnectthedevicefromtheelectricalpowersupplybeforeopeningthedevice.
Useavoltmetertoverifythatthedeviceisnotunderpower!
3. Besurethatthepowersupplycannotbereconnectedwithoutnotice.
WARNING
Dangertolifefromstrongmagneticfields!
Strongmagneticfieldsmaycauseseriousinjuriesordeathandsignificantdamageto
property.
1. Persons fitted with heart pacemakers must be kept away from the device. The
functionalityoftheheartpacemakercouldbecompromised.
2. Persons with metal implants must be kept away from the device. Implants may
heatuporbesubjecttomagneticattraction.
3. Ferromagneticmaterials,tools,and electromagnetsmustbe kept awayfromthe
magneticsource.Suchmaterialscouldbesubjecttomagneticattractionandmay
flyaroundtheroom,injuringorkillingpeople.Minimumdistance3meters.
4. Remove magnetic items (jewelry, watches, pens etc.) before carrying out
maintenancework.
5. Keep electronic equipment away from the magnetic source. Such equipment
couldbedamaged.
6. Keepstoragemedia,creditcardsetc.awayfromthemagneticsource.Datacould
beerased.
Note:Thedevicedoesnotproduceahighmagneticfield.

Table of Contents

Related product manuals