Safety
Z33081_00_001 21
DANGER
DangerofSuffocationfromHydrogenGas
If excessive amounts of H2 gas is released in an enclosed space, hydrogen will
decreasetheamountofavailableoxygenandmaycausesuffocation.
1. The device should be installed in a sufficiently spacious, well ventilated room
equippedwithapermanentH2alarmgassensor.
2. The user is referred to MSDS safety data sheets for hydrogen gas for further
information.
WARNING
Dangerofinjuryfromelectricalshock!
Alifethreateningshockmayresultwhenthehousingisopenduringoperation.
1. Onlyqualifiedpersonnelshouldopenthehousing.
2. Disconnectthedevicefromtheelectricalpowersupplybeforeopeningthedevice.
Useavoltmetertoverifythatthedeviceisnotunderpower!
3. Besurethatthepowersupplycannotbereconnectedwithoutnotice.
WARNING
Dangertolifefromstrongmagneticfields!
Strongmagneticfieldsmaycauseseriousinjuriesordeathandsignificantdamageto
property.
1. Persons fitted with heart pacemakers must be kept away from the device. The
functionalityoftheheartpacemakercouldbecompromised.
2. Persons with metal implants must be kept away from the device. Implants may
heatuporbesubjecttomagneticattraction.
3. Ferromagneticmaterials,tools,and electromagnetsmustbe kept awayfromthe
magneticsource.Suchmaterialscouldbesubjecttomagneticattractionandmay
flyaroundtheroom,injuringorkillingpeople.Minimumdistance3meters.
4. Remove magnetic items (jewelry, watches, pens etc.) before carrying out
maintenancework.
5. Keep electronic equipment away from the magnetic source. Such equipment
couldbedamaged.
6. Keepstoragemedia,creditcardsetc.awayfromthemagneticsource.Datacould
beerased.
Note:Thedevicedoesnotproduceahighmagneticfield.