EasyManua.ls Logo

Bruker BPHG 90 - Ethernet Connection; USB Connection; Mains Power Connection

Bruker BPHG 90
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation
Z33081_00_001 39
6.2.3 EthernetConnection
TheBPHG 90 can optionally beconnected viaan Ethernetlink toan externalcomputer.
ConnecttheEthernetcableontherearpanel(labeledETH)toafreeEthernetportonthe
externalcomputer,usingastandardRJ45Ethernetcable.
6.2.4 USBConnection
The generator has an USB port on rear panel for updating the firmware. This port is
reservedforusebytrainedservicepersonnelonly.
6.2.5 MainsPowerConnection
TheBPHG90isconnectedtothemainsusinga3wiregroundedpowercordratedfor600
W. Plug the power cord into the rear power connector and connect the other end to the
walloutlet.TheBPHG90cannowbeswitchedon.Therearpowerswitchonthesocket
mustbeinONposition.
ThelinesocketcontainstwoT4A/250Vfusesinasmalldrawer.Ifthefusesneedtobe
replaced,alwaysusethesametypeandrating.
TheredpowerswitchonthefrontisusedtoswitchtheinstrumentONandOFF.Aredlight
inthefrontpowerswitchindicatesthatlinevoltageispresent.

Table of Contents

Related product manuals