Operation
42 Z33081_00_001
7.1 GettingStarted
1. Open the H
2
gas supply valve and ensure that the hydrogen inlet pressure is in the
correctrange:6to10bar.
2. Switch on the generatorusing the mains power switchon thefront panel (seethe next
figure).
NexttothepoweronswitchonthefrontpaneloftheBPHG90aretwoLEDsthatprovide
basicstatusinformation.
The green Ready LED indicates normal operation of the BPHG 90. During the cooling
downoftheconversionchamber,itblinks,andwhenthepresetconversiontemperatureis
reached,itwillstaypermanentlyON.
TheredErrorLEDindicatesthatanerrorconditionispresent(e.g.alowcoolantlevel,low
inputpressure,etc.).
Figure7.3:PowerSwitchandPanelLED's