EasyManua.ls Logo

Bruker BPHG 90 - System Check Window; Stand­by Mode; Figure 7.4 System Check with Low Coolant Level Warning; Figure 7.5 Hydrogen Gas Sensor Warming up

Bruker BPHG 90
78 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
Z33081_00_001 43
7.1.1 SystemCheckWindow
When the BPHG 90 is switched on, the instrument briefly shows a welcome screen and
thenperformsapreoperationSystemCheckofthemaincomponents.Attheendofthe
systemcheck,thesystemdisplaystheSystemCheckwindow.Ifanyofthetestsdoesnot
pass,thefailedtestresultishighlightedin (seethefigurebelow)andoperationofthe
instrumentisnotpossibleuntiltheuserhastakenremedialaction.
Figure7.4:SystemCheckwithLowCoolantLevelWarning
AspecialcaseisthetestforthehydrogensensorthatmeasurestheH
2
concentrationinair
insidetheinstrument.Thissensorneedsaboutaminutetowarmupandthisisindicated
byamessagein (seethefigurebelow).
Whenalltestspass,theSystemCheckwindowdisappearsafterafewsecondsandthe
systementersStandbymode(seesectionStand‐by Mode [}43]).
Figure7.5:HydrogenGasSensorWarmingUp
7.1.2 StandbyMode
Once the instrument has successfully passed the initial automatic system check, it goes
intoSTANDBYmode.First,aconfirmationscreenisdisplayed(seethenextfigure),asking
theusertoverifythattheexhausthydrogengastubeissafelyroutedoutsidetheroomor
toafumecupboard.

Table of Contents

Related product manuals