Operation
44 Z33081_00_001
The BPHG 90 is delivered with 10 meter of flexible exhaust plastic tubing, which is
connected to the rear of the instrument and routed, for example to a fume cupboard or
outsidetheroom(seesectionExhaust Line Connection [}38]).
Figure7.6:ExhaustGasWarningMessage
WARNING
Hydrogenisaveryflammablegas.
Theexhausthydrogen gas mustbesafely routed outsidetheroom usingthetubing
supplied.
Oncethesafetycheckmessagehasbeenconfirmed,thesystemswitchestotheStandby
screen in which the user can select between the three main operation modes of the
instrument(seethenextfigure):Start,Purge,andWarmUp.
Inaddition,Settingsgivestheuseraccessto various operationalinformationandsetup
parameters.