Safety
Z33081_00_001 25
3.7.3 MechanicalDangers
CAUTION
Accidenthazardfrommovementofmechanicalparts!
Thefingersorhandmaybepinchedduetomovementofmechanicalparts.
1. Shutoffthedevicebeforeaccessing.
CAUTION
Accidentandmaterialdamagehazardfromfallingobjects!
Equipmentmayfalldownduringassembly,retrofitting,ordismantling.Thismayresult
inpersonalinjuryorequipmentdamage.
1. Ifnecessary,assemble/disassemblethedeviceinmultipleparts.
2. Useaplatformwithrailingsinsteadofaladdertoreachtheassemblyarea.
3. Avoid working over the head. When this can not be avoided, wear a protective
hardhat.
4. Followthemountinginstructionsintheinstallationmanual.
3.7.4 DangersfromMagneticFields
WARNING
Dangertolifefromstrongmagneticfields!
Strongmagneticfieldsmaycauseseriousinjuriesordeathandsignificantdamageto
property.
1. Persons fitted with heart pacemakers must be kept away from the device. The
functionalityoftheheartpacemakercouldbecompromised.
2. Persons with metal implants must be kept away from the device. Implants may
heatuporbesubjecttomagneticattraction.
3. Ferromagneticmaterials,tools,and electromagnetsmustbe kept awayfromthe
magneticsource.Suchmaterialscouldbesubjecttomagneticattractionandmay
flyaroundtheroom,injuringorkillingpeople.Minimumdistance3meters.
4. Remove magnetic items (jewelry, watches, pens etc.) before carrying out
maintenancework.
5. Keep electronic equipment away from the magnetic source. Such equipment
couldbedamaged.
6. Keepstoragemedia,creditcardsetc.awayfromthemagneticsource.Datacould
beerased.
Themagneticfieldofthedevicedoesnotcauseanypersonalinjuriesorpropertydamage.
Forfurtherinformationseethemanualofthemagnetused.