Operation
46 Z33081_00_001
7.2.1 AdjustableParameters
In the PHG > Settings window, the user has access to the following operational
parameters:
• Conversiontemperaturesetpoint(36...40K),default36K.
• Batchdeliverytargetpressure(0...6bar),default3bar.
• Fillingcontainerdeliverytimeout(0...60min),default60min.
Figure7.9:UserSettingsMenuwithAdjustableParameters
In order to change a parameter value, tap the parameter field. This will bring up a
numericalkeypadtoenterthenewvalue.Theuseraccessibleparametershavealimited
valuerangeandenteringvaluesoutsidethisrangewillcausethenearestvalidvaluetobe
loaded.Theparametervaluesarestoredinanonvolatilememoryanduserdefinedvalues
will survive a powerdown of the instrument. Factory preset values can be restored by
tappingtheDefaultbuttonnexttothevaluefield.
7.2.2 OperatingInformation
The BPHG 90 has builtin timers for monitoring the operation of maintenance critical
components. The tab Operating Information brings up an information window with the
values of the operating times of these components and of the main operational
parameters.
Figure7.10:OperationInformationScreen