110
ZTKS0156 / Z31820 / Rev.: 03
Troubleshooting
6.3.3 Procedure in case of an alarm signal
If the temperature of the radiation shield exceeds the set maximum allowed temperature,
an alarm will warn the operator (audible from the External Temperature Control Unit,
visible from the External Temperature Control Unit and the console via MICS).
Follow this procedure to fix the issue before contacting Bruker Service.
Note:
Date and time.
Temperature given on the display of the External Temperature Control Unit.
Turn the alarm off (see supplied manual of the External Temperature Control Unit).
Which errors or warnings appear?
What are the readings of the pressure gauges, run time counter etc. of the
compressor?
Which parts seems to be ok/not ok?
How did the failure occur and what happened before failure?
Figure 6.4: Changing the Helium level sensor - step 3
12. Fix the black nut (1).
13. Calibrate the helium level sensor in the
TopSpin Service Web. Refer to the manual
of TopSpin.