Troubleshooting
6.2 Problem
6.2.1 During Transportation
6.2.2 During Assembling
Continued on next page
Indicator Possible reason Solution By
Tilt Watch / Shock
Watch activated.
Careless transporta-
tion.
1. Accept delivery with
reservation.
2. Remark the extent of
damage in the transport
documents.
3. Start complaint process.
Approved
Customer
Personnel
Visible damage. Careless transporta-
tion.
1. Accept delivery with
reservation.
2. Remark the extent of
damage in the transport
documents.
3. Start complaint process.
Approved
Customer
Personnel
Indicator Possible reason Solution By
Ceiling height too
low for assembling
on magnet stand.
Site does not meet
the required condi-
tions.
Choose another site that
meets the required condi-
tions.
Bruker Service
Ceiling height too
low for inserting the
helium level sensor.
Site does not meet
the required condi-
tions.
Insert the helium level sensor
before mounting the magnet
stand.
Bruker Service
Helium bore tube
and radiation shield
are not concentric.
Alignment is not
correct.
Check fixation of the
alignment rods.
Bruker Service