BTL-5000 SERIES
USER'S MANUAL
page 23 of 56
2.9 THERAPY PRINTING AND SAVING
Pressing the print, save button allows you to make several choices. Simply complete the form by entering the
required data filed as shown on the screens below.
2.9.1 Save Therapy
You can save your therapy after setting therapy parameters from the therapy parameters screen –see Chapter 2.5.6
Therapy Parameters Screen – Ergonomic, Standard and Expert Mode.
The following information is saved with each therapy:
Electrotherapy:
• All parameters of currents (pulse length, pause length, modulation, etc.)
• Therapy time
• Polarity
• Output mode (current / voltage)*
Ultrasound therapy:
• All therapy parameters (for example, ultrasound frequency, duty factor - DF, pulse frequency, etc.)
• Therapy time
• Intensity
Combined therapies electro + ultrasound:
• All electrotherapy parameters (pulse length, pause length, modulation, etc.)
• All ultrasound therapy parameters (ultrasound frequency, duty factor - DF, pulse frequency, etc.)
• Therapy time
• Polarity of electrotherapy output
• Electrotherapy output mode (current / voltage)
• Intensity of ultrasound*
Laser therapy:
• All therapy parameters (frequency, course of signal, etc.)
• Irradiated area
• Dosage
* output intensity can be entered into a comment (e.g. at threshold motor level)
When saving therapy, enter:
• Name of diagnosis (therapy) – to be displayed in the list of diagnoses - the diag (8) button
• Name of program – to be displayed in the list of programs - the prog (9) button
• Description, additional information – to be displayed in both lists
The unit suggests the lowest available number (from the range of 8000-8999) and adds the letter of the corresponding
generator (E for electrotherapy and combined therapies, U for ultrasound therapies, and L for laser therapies).
2.9.2 Save Therapy and Add It to the Patient Data
The therapy is saved as described above and assigned to the patient in their list of therapies.