BTL-5000 SERIES
USER'S MANUAL
page 31 of 56
3.2.8 Operation Mode
Selects one of the three modes, see Chapter 2.5.6 Therapy Parameters Screen – Ergonomic, Standard and
Expert Mode.
Factory-preset is ergonomic mode.
3.2.9 Touch Panel Calibration
If the buttons on the touch screen do not react when pressed, the touch screen needs calibration. Calibration values
are displayed on the screen and the soft touch stylus is used to make adjustments to the sensitivity of the buttons.
Press esc to stop calibration. To verify touch screen adjustments, use the 'touch panel function test' function.
Press ‘ESC’ to stop calibration.
To verify touch screen adjustments, use the "TOUCH PANEL FUNCTION TEST".
3.2.10 User Options
Here, you can set:
• direction of cursor movement when using the select (14) control
• listing of therapies and some other menu options (in ascending or descending alphabetical order)
• location of the tab bar (up / down)
• speaker volume
3.2.11 Way of Operation
3.2.11.1 New Operation Mode
Use this option for the first series BTL-5000 units manufactured in 2001 and early in the year 2002. These units have
the knobs (17) and (18) marked as time (17) and start / stop (18). For these units, select the option new operation
mode = no (after the new version of firmware is loaded).
New units are factory pre-set to new operation mode (new operation mode = yes).
3.2.11.2 End of Therapy – Setting Zero Intensity and Time
Values
After the end of therapy, you can have displayed either zero values of intensity and time or the intensity and time
values of the last performed therapy.
3.2.12 Network Setting
This option is applicable only if the device is connected in network. Here it is possible to set the name by which the
device will report in the network. It is possible to set if the device shall automatically login in the network after startup
or if it will be logged in only after pressing a button. This item also provides test of functionality of the network and
information about all devices connected in the network.
For operation of the devices in the network it is necessary to have purchased the software package for the master
computer; in addition, the network must be professionally installed by an authorized distributor's representative.
Detailed information about network interconnection and control is included in the software package.
3.2.13 Information about the Unit
Displays info about the unit (serial number, firmware version, etc.).It also contains the information till when the device
will work – so called "device validity". If the functioning of the device is temporary, this item contains the information
until which date the device will be fully functional.
3.2.14 Service Functions
3.2.14.1 Repair of Files
Checks the file system in the unit and repairs possible errors -- deletes empty files, etc. Recommended for use in case
of lack of memory, if the unit rejects to save data, or if you think that some data have been lost.