8200-M585-O1 Rev D 210 Installation & Technical 34
SETUP AND CALIBRATION, CONT.
int= (Interval Setting)
Press the ENTER key to show the current value.
If LFt = 1 (Legal For Trade = YES), using the numeric keys enter the new setting, then
press the ENTER key to save it. Allowable values are: 1, 2, 5, 10, 20 or 50.
If LFt=0 (Legal For Trade = NO), using the numeric keys enter the new setting, then press
the ENTER key to save it. Allowable values are: 1 through 99.
In either case, if the setting displayed is acceptable, press the ENTER key again it.
dPP= (Decimal Point Setting)
Press the ENTER key to show the current value. If the setting displayed is acceptable, press
the ENTER key again to save it. Otherwise, using the numeric keys enter the new setting,
then press the ENTER key to save it. Allowable values are: 0, 1, 2 or 3.
0 = X X X X X X 2 = X X X X . X X
1= X X X X X . X 3 = X X X . X X X
CAP= (Capacity)
Press the ENTER key to show the current value. If the setting displayed is acceptable, press
the ENTER key again to save it. Otherwise, using the numeric keys enter the new setting,
then press the ENTER key to save it. Allowable values are: 1 through 999,999.
NOTE: Capacity cannot exceed 999,999.
Unit2= (Weighing Unit 2)
Press the ENTER key to show the current value. If the setting displayed is acceptable, press
the ENTER key again to save it. Otherwise, using the numeric keys enter the new setting,
then press the ENTER key to save it. Allowable values are:
0 = none 4 = oz (ounces)
1= tn (tons) 5 = kg (kilograms)
2= g (grams) 6 = tonnes (metric tons)
3 = lb (pounds) 7 = lb/oz (pounds/ounces)
NOTE: The selection for Unit2 can
not be the same as Unit1. In addition,
dependent upon the selection for Unit1 and the interval and decimal point settings,
not all unit combinations are available.
trA= (Zero Tracking Range)
Press the ENTER key to show the current value. If the setting displayed is acceptable, press
the ENTER key again to save it. Otherwise, using the numeric keys enter the new setting,
then press the ENTER key to save it. Allowable values are: 0 (disables Zero Tracking), .5, or 1
through 9.
trL= (4% Zero Range)
Press the ENTER key to show the current value. If the setting displayed is acceptable, press
the ENTER key again to save it. Otherwise, using the numeric keys, 0/NO or 1/YES, enter the
new setting, then press the ENTER key to save it.
trL=1 (Yes) trL=0 (No)
4% of scale capacity Full capacity (no limit)