8200-M585-O1 Rev D 210 Installation & Technical 35
SETUP AND CALIBRATION, CONT.
PUO= (Power-Up Zero Feature)
Press the ENTER key to show the current value. If the setting displayed is acceptable, press
the ENTER key again to save it. Otherwise, using the numeric keys, 0/NO or 1/YES, enter the
new setting, then press the ENTER key to save it.
PUO=1 (Yes) PUO=0 (No)
Automatic Re-Zero on Power-Up No Re-Zero on Power-Up
td = (12 or 24 Time Format)
Press the ENTER key to show the current value. If the setting displayed is acceptable, press
the ENTER key again to save it. Otherwise, use the numeric keys to select the format (12 or
24 hour) of the Model 210 clock operation, then press the ENTER key to save it. Note that in
the 24 hour format, 12 is added to all times after noon, i.e. 3 PM would be 1500.
td = 12 td = 24
12 hour clock (3PM displays 3:00) 24 hour clock (3PM displays 15:00)
SLEEP= (Sleep Mode Feature)
The Sleep Mode feature conserves power when the indicator remains unused for a selected
period of time. With the feature enabled, the load cell excitation will be reduced and the
display will be blank.
Press the ENTER key to show the current status of this feature. If a number other than 0 is
shown, this feature is selected and the number shown corresponds to the number of minutes
of a stable zero weight reading before the indicator enters the sleep mode. If the setting
displayed is acceptable, press the ENTER key again to save it. Otherwise, use the numeric
keys to enter a new value (0 to 10) then press the ENTER key to store the new setting. Note
that entry of a 0 disables this feature.
A oFF= (Auto Shutoff)
The Automatic Shutoff feature will automatically turn the indicator off (when it is not in use)
after a predetermined period of inactivity to conserve power. To turn the indicator back on you
must press the ON/OFF key.
Press the ENTER key to show the current status for this feature. A number other than 0
indicates that the auto shutoff feature is enabled and the displayed number corresponds to the
number of minutes of stable weight displayed before the indicator is turned off automatically.
Note that a 0 indicates the feature has been turned off. If the setting displayed is acceptable,
press the ENTER key again to save it. Otherwise, use the numeric keys to enter a new value
(0 to 10) then press the ENTER key to store the new setting.
CLtAr= (Clear Tare)
The Clear Tare feature allows the indicator to clear the Stored Tare weight when the Net
weight goes below a value greater than 1/2 the stored tare weight or goes below zero (a
negative net weight after display of a positive net weight). With this feature enabled, the
operator must re-set the tare after completion of a transaction when the load (container plus
item) is removed from the scale.
Press the ENTER key to show the current value. If the setting displayed is acceptable, press
the ENTER key again to save it. Otherwise, using the numeric keys, 0/NO or 1/YES, enter the
new setting, then press the ENTER key to save it.
CLtAr=1 (Yes) CLtAr=0 (No)
Automatically clears Stored Tare
when Net weight goes below zero
Stored Tare is not cleared when
Net weight goes below zero