64
User’s Manual
CellaVision
®
DM1200
(Cont'd)
Including and Excluding Slides in/from the Reported Result
Signed slides are automatically included in the reported result.To include or
exclude a slide, activate/deactivate the checkbox next to the Slide ID.
When excluding slides, a dialog is shown where you may write a comment
explaining the exclusion.
Note! You can only include signed slides.
Note! You should not merge slides if one or more slides have a "confirm cell
counter result" flag. If you do, the calculated result in the reported result column
will be displayed incorrectly.
Changing the Reported Result
It is possible to change the RBC results and the PLT concentration for the order, if
reported as a level. Changeable results are written in bold text.
1. Click on the result to change.
2. Change the value in the dialog.
3. Click OK.
Note! If you include or exclude a slide, all manually changed values will be
replaced by new, automatically calculated values.
(Cont'd)