148
User’s Manual
CellaVision
®
DM1200
Appendix D — Recommended Workflow
Laboratories handle their peripheral blood differential counts in different ways
making it hard to suggest one general workflow suitable for all laboratories. See
flowcharts of the three workflows below.
Recommended Settings (see 9.7 Adjusting Report/Sign Settings)
1. Single slide differentials
Pre-fill password: Enabled
Sign order when signing the slide: Enabled
Send order to LIS when signed: Enabled
Review the cells in the WBC Galleries.
2. Confirmation of cell counter results - Quickly scanned slides for verification of
cell counter results.
Pre-fill password: Enabled
Sign order when signing the slide: Enabled
Send order to LIS when signed: Enabled
Review the analysis type you want to confirm and then click Confirm Cell Counter
Results.
When using the Confirm Cell Counter Result for any analysis type (WBC, RBC, or
PLT) the normal sign slide checks (cells in the Unidentified class, all cells
reviewed etc.) are disabled.
Note! When using Confirm Cell Counter Results the results sent to the LIS are: a
WBC confirmation flag, a RBC confirmation flag and/or a PLT confirmation flag
For a slide it is possible to report the WBC result and a RBC confirmation flag and
a PLT result or any other combination.
3. Duplicate slides - An order consists of two slides. Two persons sign one slide
each.
Pre-fill password: Enabled
Sign order when signing the slide: Disabled
Send order to LIS when signed: Enabled
Review the cells in the WBC Galleries.
Confirm Cell Counter Results