6
Safety
In addition, adequate ventilation should be provided for preparation of reagents and solvents for use on the
system. All solvent bottles and the Waste Reservoir should be placed into proper secondary containers to minimize
the risk of exposure.
Operating Precautions
The Liberty Blue must be grounded. In the event of an electrical short circuit, grounding reduces the risk of electric
shock by providing an escape wire for electric current. This instrument is equipped with a cord having a grounding
wire with a grounding plug. The plug must be plugged into an outlet that is properly installed and grounded.
Consult a qualied electrician or service technician if the grounding instructions are not completely understood or
if doubt exists as to whether the instrument is properly grounded. If it is necessary to use an extension cord, use
only a 3-wire extension cord that has a 3-blade grounding plug and a 3-slot receptacle that will accept the plug
from the instrument. The marked rating of the extension cord must be equal to or greater than the electrical rating
of the instrument.
Do not replace detachable main supply cord with inadequately rated cords.
The possibility of instrument-induced electromagnetic interference (EMI) is minimal if the instrument is operated as
outlined in this manual. The instrument should not be placed close to any electrical device susceptible to EMI.
This instrument utilizes high voltages and microwave radiation. Only those trained in repair and maintenance of
high voltage and microwave power systems should perform instrument service and repair.
This instrument complies with United States Code of Federal Regulations 21 CFR Part 1030.10 (C) for microwave
leakage. This instrument complies with FCC Requirements in the United States Code of Federal Regulations
(47CFR Part 18) – Industrial, Scientic and Medical (ISM) Equipment – emissions requirements.
WARNING
Only accessories which meet the manufacturer’s specications shall be used.
Do not attempt to lift or move the Liberty Blue system. Contact CEM Corporation if the Liberty Blue system
needs to be moved.
If instrument is used in a manner not specied by CEM Corporation in this manual, then the protection provided by
the equipment may be impaired.

Table of Contents

Other manuals for CEM Liberty Blue

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CEM Liberty Blue and is the answer not in the manual?

CEM Liberty Blue Specifications

General IconGeneral
BrandCEM
ModelLiberty Blue
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals