EasyManuals Logo

CEM Liberty Blue User Manual

CEM Liberty Blue
63 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #49 background imageLoading...
Page #49 background image
45
Maintenance of the Liberty Blue
4. Enter the volume to be used for calibration, then click OK. Typically, 5 mL of reagent are used for calibration.
5. The selected reagent will be added to the reaction vessel. Once the addition is nished, measure the volume
delivered to the reaction vessel using the graduations on the calibration stand.
6. Enter the volume into the Delivered Volume box.
7. Click OK to drain the reaction vessel. The software will then calculate the new ow rate. Click OK to save the
new calibration values.
8. Once the new rate has been saved, click OK to complete the operation.
9. If the new timed addition ow rate is less than 1.0 mL/s, replace the position’s lter with PN 551000 and
recalibrate. If the ow rate is still below 1.0 mL/s, contact CEM.
Reaction Vessel Drain Test
1. In the Liberty Blue application software, open the Maintenance screen.
2. Under Addition, select the Add Main Wash (MS2) operation.
3. On the Parameters tab, double click the value for Volume, enter 10 in the pop-up, then click OK.
4. Click the Run Operation button to execute the operation. This will add 10 mL of solvent to the vessel.
5. Click the Drain button at the bottom of the Maintenance screen.
6. The liquid should drain within 10 seconds. If the liquid requires more than 10 seconds to fully drain, replace the
entire vessel (PN 167260).
Verification of Waste Sensor Operation
The Waste Sensor is a oat sensor located on the top of the waste bottle. It is connected to the Liberty Blue
through a 9-pin connector cable plugged into the left side of the Liberty Blue module. Periodically, operation of the
Waste Sensor should be veried.
WARNING
Handle all waste under a fume hood, and wear suitable protective clothing. Dispose of all waste in
accordance with all applicable local, state, and federal health and safety recommendations.
1. Ensure the Liberty Blue is idle.
2. Remove the cap from the waste container.
3. Lift the oat sensor.

Table of Contents

Other manuals for CEM Liberty Blue

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the CEM Liberty Blue and is the answer not in the manual?

CEM Liberty Blue Specifications

General IconGeneral
BrandCEM
ModelLiberty Blue
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals